PT. MF is a company engaged in the
pharmaceutical with the end product of quality drugs.
Good drug quality can be guaranteed with the right
quality control, while the right quality control is by
checking the drug level by using High Performance
Chromatography Liquid (HPLC) machine with the
most important part is HPLC column. During the
period of January - December 2017, the HPLC column
of type C-18, 250 x 4 mm often suffered damage by
10.20% with a sigma value of 3.14 this has exceeded the
predetermined target of 2.0%. The purpose of this
research is to find factors causing damage of column,
give suggestion of repair and compare result before and
after repair by using DMAIC method so that factor
causing damage of HPLC column in factor method is
HPLC column leaching process not according to
supplier recommendation can be done that is make
HPLC column leaching procedure in accordance with
supplier recommendation, and do training related
HPLC column leaching. By applying the proposed
improvements and controls, the proposed
improvements can be effective and efficient so as to
reduce the damage of C-18, 250 x 4 mm HPLC columns
and increase the sigma value during the January - May
2018 period of 1.70% and 3, 89.
Keywords : Quality, HPLC Column Damage, DMAIC, Sigma Value.